High-precision GAFCHROMIC EBT film-based absolute clinical dosimetry using a standard flatbed scanner without the use
Heeteak Chung1, Bart Lynch, Sanjiv Samant
1Department of Nuclear and Radiological Engineering, University of Florida, Gainesville Florida, USA. heechy@ufl.edu
Abstract:
To report a study of the use of GAFCHROMIC EBT radiochromic film (RCF) digitized with a commercially available flatbed document scanner for accurate and reliable all-purpose two-dimensional (2D) absolute dosimetry within a clinical environment. We used a simplified methodology that yields high-precision dosimetry measurements without significant postirradiation correction. The Epson Expression 1680 Professional scanner and the Epson Expression 10000XL scanner were used to digitize the films. Both scanners were retrofitted with light-diffusing glass to minimize the effects of Newton rings. Known doses were delivered to calibration films. Flat and wedge fields were irradiated with variable depth of solid water and 5 cm back scatter solid water. No particular scanner nonuniformity effect corrections or significant post-scan image processing were carried out. The profiles were compared with CC04 ionization chamber profiles. The depth dose distribution was measured at a source-to-surface distance (SSD) of 100 cm with a field size of 10 x 10 cm2. Additionally, 22 IMRT fields were measured and evaluated using gamma index analysis. The overall accuracy of RCF with respect to CC04 was found to be 2%-4%. The overall accuracy of RCF was determined using the absolute mean of difference for all flat and wedge field profiles. For clinical IMRT fields, both scanners showed an overall gamma index passing rate greater than 90%. This work demonstrated that EBT films, in conjunction with a commercially available flatbed scanner, can be used as an accurate and precise absolute dosimeter. Both scanners showed that no significant scanner nonuniformity correction is necessary for accurate absolute dosimetry using the EBT films for field sizes smaller than or equal to 15 x 15 cm2.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
10:33Expedited Radiation Biodosimetry by Automated Dicentric Chromosome Identification (ADCI) and Dose Estimation
Published on: September 4, 2017