Reliability of Artificial Intelligence-Based Cone Beam Computed Tomography Integration with Digital Dental Images
Published on: February 23, 2024
Effects of tube current on cone-beam computerized tomography image quality for presurgical implant planning in vitro
Jaideep Sur1, Kenji Seki, Hiroshi Koizumi
1Department of Radiology, Showa University School of Dentistry, Tokyo, Japan.
Objective:
The purpose of this study was to evaluate the effect of tube current reduction on the quality of cone-beam computerized tomography (CBCT) images of the maxilla and mandible for presurgical implant planning.
Study Design:
Six cadaver heads were imaged with a CBCT (3D Accuitomo) operated at 80 kVp and variable tube current (8, 4, 2, or 1 mA) in full-scan mode. For the posterior region of maxilla and mandible, half-scans were also performed at similar settings. The images were evaluated independently by 5 oral radiologists for 15 anatomic landmarks regarding presurgical implant planning. The quality of images were compared with their respective reference images at 8 mA and ranked on a 4-point rating scale as excellent, good, fair, or nondiagnostic. The scores of all observers were averaged for each landmark at every exposure condition, and Bonferroni test (P<.05) was performed.
Results:
The 4 mA images at full-scan mode could visualize each landmark of maxilla and mandible and were evaluated to be the same or almost equivalent in quality as the 8 mA images. Even 2 mA images in full-scan mode and 4 mA in half-scan mode could be used for implant planning. The 1 mA images were unacceptable owing to the substantial degradation in image quality.
Conclusion:
Significant dose reduction can be achieved by reducing tube current without substantial loss of image quality for presurgical implant planning in CBCT.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...