Accuracy and precision of buccal pulse oximetry
Marla J De Jong1, Joseph Schmelz, Karen Evers
1TriService Nursing Research Program, Uniformed Services University of the Health Sciences, Bethesda, Maryland 20814, USA. mdejong@aol.com
Objective:
We sought to describe the accuracy and precision of buccal pulse oximetry (SbpO(2)) compared with arterial oxygen saturation (SaO(2)) and pulse oximetry (SpO(2)) in healthy adults at normoxemia and under 3 induced hypoxemic conditions.
Methods:
In this prospective, correlational study, SbpO(2), SaO(2), and SpO(2) values were recorded at normoxemia and at three hypoxemic conditions (SpO(2)=90%, 80%, and 70%) for 53 healthy, nonsmoking adults who were without cardiac or pulmonary disease, baseline hypoxemia, peripheral edema, dyshemoglobinemia, and fever. Bland-Altman analyses were used to assess agreement and precision between SbpO(2) and SaO(2) measures and between SbpO(2) and SpO(2) measures. Data were adjusted to account for a lag time between buccal and finger sites.
Results:
When comparing SbpO(2) and SaO(2) values, mean differences of -1.8%, .3%, 2.4%, and 2.6% were evident at the normoxemia, 90%, 80%, and 70% levels, respectively. When comparing SbpO(2) and SpO(2) values, the mean differences were -1.4%, .1%, 3.3%, and 4.7% at the normoxemia, 90%, 80%, and 70% levels, respectively. The SbpO(2) and SaO(2) values met a priori precision criteria (1.6%; 95% confidence limit, -4.9% to 1.3%) at normoxemia. The SbpO(2) and SpO(2) values met precision criteria at normoxemia (1.5%; 95% confidence limit, -4.4% to 1.5%) and 90% (1.9%; 95% confidence limit, -3.6% to 3.8%) conditions, but exceeded precision criteria at the other tested conditions. On average, SpO(2) lagged 21 seconds behind SbpO(2).
Conclusion:
Buccal oximetry is an inaccurate and imprecise method of assessing SpO(2) when oxygen saturation is <90%. The divergence between SbpO(2) and both SaO(2) or SpO(2) values increased as hypoxemia worsened. The buccal method overestimated oxygen saturation in proportion to the degree of hypoxemia. Such overestimates may lead nurses to conclude falsely that a patient's arterial oxygen saturation is acceptable when further assessment or intervention is warranted.
More Related Videos
07:51Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Special considerations while measuring pulse
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...