Meniscal repair using large diameter horizontal sutures increases fixation strength: an in vitro study
Yavuz Kocabey1, Omer Taşer, Onur Hapa
1Department of Orthopaedics and Traumatalogy, Kocaeli Acıbadem Hospital, Kocaeli, Turkey. drkocabey@yahoo.com
Summary
Larger diameter sutures provide superior mechanical fixation for meniscal repair. This study found that #2 Fiberwire suture demonstrated greater load at failure and stiffness compared to #2-0 Fiberwire suture in bovine meniscal models.
Area of Science:
- Orthopedic Surgery
- Biomechanical Engineering
- Sports Medicine
Background:
- Meniscal tears are common knee injuries requiring surgical repair.
- The choice of suture material and diameter significantly impacts the mechanical stability of meniscal repairs.
- Understanding the biomechanical properties of different suture configurations is crucial for optimizing surgical outcomes.
Purpose of the Study:
- To compare the mechanical characteristics of meniscal repair fixation using horizontal sutures of two different diameters (#2-0 and #2 Fiberwire).
- To evaluate suture performance under submaximal cyclic loading and load-to-failure conditions.
Main Methods:
- Bovine medial menisci were used, with a 2-cm longitudinal incision created.
- Repairs were performed using either #2-0 (Group 1) or #2 (Group 2) Fiberwire suture.
- Specimens underwent cyclic testing (5-50 N at 1 Hz for 500 cycles) and load-to-failure testing (5 mm/min crosshead speed).
Main Results:
- Group 2 (larger diameter suture) exhibited significantly greater load at failure (132.1 ± 54.4 N) compared to Group 1 (91.9 ± 26.2 N) (P = 0.02).
- Group 2 also showed significantly greater stiffness (47.1 ± 8.3 N/mm) than Group 1 (38.5 ± 10.2 N/mm) (P = 0.03).
- Failure mode in all specimens was suture pull-through, indicating suture strength and tissue interface limitations.
Conclusions:
- Larger diameter sutures (#2 Fiberwire) provide superior mechanical fixation for meniscal repairs compared to smaller diameter sutures (#2-0 Fiberwire).
- The findings suggest that utilizing larger diameter sutures in horizontal meniscal repair techniques can enhance repair strength and stability.