Analysis of leucocyte populations with the Coulter S-Plus STKR as a screening tool for haematological abnormalities

G d'Onofrio1, A M Salvati, P Berti

  • 1Servizio di Ematologia, Università Cattolica del Sacro Cuore, Roma, Italia.

In two institutions at Rome and Florence we evaluated the clinical sensitivity of two Coulter STKR systems using the NCCLS standard H20-T for leucocyte differential count in a patient population with high prevalence of haematologic abnormalities. Reference ranges of normal leucocytes were obtained on 278 adult subjects. On a population of 455 patient specimens, 200 specimens (44%) were flagged by the STKR because of a distributional abnormality, and 122 (27%) because of a morphological abnormality. Percentage of subtotal agreements between the STKR and the reference manual differential count was 85.4%, with 67.5% full and 20.9% partial agreements. Eight specimens that showed a morphological abnormality with the reference manual differential count were classified as normal by the STKR, with a false normal rate of 6.6%. Analysis of the STKR performance for morphological abnormalities showed acceptable sensitivity (82.0%) and rather low specificity (71.5%), low predictive value of positive results (51.3), high predictive value of negative results (91.5%) and efficiency of 74.3%. The main problems of the STKR differential count were a high rate of false monocyte count, and the misidentification of eosinophilias and low-concentration abnormal cells.