Electrically assisted solid-phase microextraction combined with liquid chromatography-mass spectrometry for
1Department of Chemistry, National Chung Hsing University, Taichung, Taiwan.
Abstract:
A novel method for electrically assisted microextraction coupled to liquid chromatography-mass spectrometry was evaluated for determination of trace levels of parathion in water. A pencil lead electrode was used in a di-electrode system to extract parathion onto the electrode surface with a reductive potential applied. The optimum extraction conditions were found to be a potential of -600 mV for 60s in pH 2 phosphate buffer solution. The parathion was desorbed statically for 1 min and dynamically for 3 min in the commercial SPME-HPLC desorption chamber, then analyzed with LC-APCI-MS/MS. The detection limit (LOD) for parathion in water was found to be 0.3 ng/mL. The proposed technique was demonstrated to be fast, sensitive and not require a solvent sample pretreatment.
More Related Videos
Related Concept Videos
High-Performance Liquid Chromatography: Types of Detectors
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...