Detection of respiratory tumour motion using intrinsic list mode-driven gating in positron emission tomography
Florian Büther1, Iris Ernst, Mohammad Dawood
1European Institute for Molecular Imaging, University of Münster, Mendelstraße 11, 48149, Münster, Germany. butherf@uni-muenster.de
Purpose:
Respiratory motion of organs during PET scans is known to degrade PET image quality, potentially resulting in blurred images, attenuation artefacts and erroneous tracer quantification. List mode-based gating has been shown to reduce these pitfalls in cardiac PET. This study evaluates these intrinsic gating methods for tumour PET scans.
Methods:
A total of 34 patients with liver or lung tumours (14 liver tumours and 27 lung tumours in all) underwent a 15-min single-bed list mode PET scan of the tumour region. Of these, 15 patients (8 liver and 11 lung tumours in total) were monitored by a video camera registering a marker on the patient's abdomen, thus capturing the respiratory motion for PET gating (video method). Further gating information was deduced by dividing the list mode stream into 200-ms frames, determining the number of coincidences (sensitivity method) and computing the axial centre of mass of the measured count rates in the same frames (centre of mass method). Additionally, these list mode-based methods were evaluated using only coincidences originating from the tumour region by segmenting the tumour in sinogram space (segmented sensitivity/centre of mass method). Measured displacement of the tumours between end-expiration and end-inspiration and the increase in apparent uptake in the gated images served as a measure for the exactness of gating. To estimate the accuracy, a thorax phantom study with moved activity sources simulating small tumours was also performed.
Results:
All methods resolved the respiratory motion with varying success. The best results were seen in the segmented centre of mass method, on average leading to larger displacements and uptake values than the other methods. The simple centre of mass method performed worse in terms of displacements due to activities moving into the field of view during the respiratory cycle. Both sensitivity- and video-based methods lead to similar results.
Conclusion:
List mode-driven PET gating, especially the segmented centre of mass method, is feasible and accurate in PET scans of liver and lung tumours.
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...