Response of thermocouples interfaced to electrothermometers when immersed in 5 water bath temperatures
Blaine C Long1, Lisa S Jutte, Kenneth L Knight
1Department of Health and Human Performance, Oklahoma State University, Stillwater, USA. blaine.long@okstate.edu
Context:
Thermocouples and electrothermometers are used in therapeutic modality research. Until recently, researchers assumed that these instruments were valid and reliable.
Objective:
To examine 3 different thermocouple types in 5 degrees C, 15 degrees C, 18.4 degrees C, 25 degrees C, and 35 degrees C water baths.
Design:
Randomized controlled trial.
Setting:
Therapeutic modality laboratory.
Intervention(S):
Eighteen thermocouple leads were inserted through the wall of a foamed polystyrene cooler. The cooler was filled with water. Six thermocouples (2 of each model) were plugged into the 6 channels of the Datalogger and 6 randomly selected channels in the 2 Iso-Thermexes. A mercury thermometer was immersed into the water and was read every 10 seconds for 4 minutes during each of 6 trials. The entire process was repeated for each of 5 water bath temperatures (5 degrees C, 15 degrees C, 18.4 degrees C, 25 degrees C, 35 degrees C).
Main Outcome Measure(S):
Temperature and absolute temperature differences among 3 thermocouple types (IT-21, IT-18, PT-6) and 3 electrothermometers (Datalogger, Iso-Thermex calibrated from -50 degrees C to 50 degrees C, Iso-Thermex calibrated from -20 degrees C to 80 degrees C).
Results:
Validity and reliability were dependent on thermocouple type, electrothermometer, and water bath temperature (P < .001; modified Levene P < .05). Statistically, the IT-18 and PT-6 thermocouples were not reliable in each electrothermometer; however, these differences were not practically different from each other. The PT-6 thermocouples were more valid than the IT-18s, and both thermocouple types were more valid than the IT-21s, regardless of water bath temperature (P < .001).
Conclusions:
The validity and reliability of thermocouples interfaced to an electrothermometer under experimental conditions should be tested before data collection. We also recommend that investigators report the validity, the reliability, and the calculated uncertainty (validity + reliability) of their temperature measurements for therapeutic modalities research. With this information, investigators and clinicians will be better able to interpret and compare results and conclusions.
Related Concept Videos
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Temperature and Thermal Equilibrium
The concept of temperature has evolved from the common concepts of hot and cold. The scientific definition of temperature explains more than just our sense of hot and cold. Temperature is operationally defined as the quantity measured with a thermometer. Furthermore, temperature is...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Calorimetry
Thermometers and Temperature Scales
As many physical properties depend on temperature, the variety of thermometers is...
Hyperthermophilic Bacteria