Variety of expiratory resistance between different continuous positive airway pressure devices for preterm infants
Martin Wald1, Angela Kribs, Valerie Jeitler
1Department of Pediatrics and Adolescent Medicine, Division of General Pediatrics and Neonatology, Medical University of Vienna, Vienna, Austria. martin.wald@meduniwien.ac.at
Abstract:
Continuous positive airway pressure (CPAP) systems for preterm infants work with conventional ventilators or use a jet ventilation system. It is assumed that the most important advantage of jet-CPAP systems is a lower expiratory resistance (R(E) ). We investigated the R(E) of seven different CPAP systems. We studied two primary-care CPAP systems, three jet-CPAP generators, and two conventional CPAP devices. All devices were adjusted at 6 mbar and connected with a test lung simulating a standardized expiration volume. Maximum pressure increase during expiration was measured and maximum R(E) was calculated. In primary-care CPAP devices, the maximum R(E) of the Benveniste valve was 9.7 mbar/L/s (SD 1.2) while that of the Neopuff was 102.8 mbar/L/s (SD 7.9) (P < 0.01). In jet-CPAP devices, the R(E) of the Infant Flow was 6.8 mbar/L/s (SD 1.7), the one of the Medijet REF 1000 was 43.5 mbar/L/s (SD 1.5), and that of the Medijet REF 1010 was 36.7 mbar/L/s (SD 0.3) (P < 0.01). In conventional CPAP systems, the R(E) of the Baby Flow was 29.7 mbar/L/s (SD 1.1) and that of the Bubble CPAP was 37.1 mbar/L/s (SD 4.3) (P < 0.01). All CPAP devices created an R(E). Jet-CPAP devices did not produce lower R(E) than conventional CPAP devices.
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Mechanical Ventilation I: Indication and Settings
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs during...
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...