Related Experiment Video
Updated: Jun 11, 2026

Development and Evaluation of 3D-Printed Cardiovascular Phantoms for Interventional Planning and Training
Published on: January 18, 2021
Development of a range of anatomically realistic renal artery flow phantoms
Deirdre M King1, Michael Ring, Carmel M Moran
1School of Physics, Dublin, Ireland. king.deirdre@mayo.edu
Abstract:
Computer-aided modelling techniques were used to generate a range of anatomically realistic phantoms of the renal artery from medical images of a 64-slice CT data set acquired from a healthy volunteer. From these data, models of a normal healthy renal artery and diseased renal arteries with 30%, 50%, 70% and 85% stenoses were generated. Investment casting techniques and a low melting point alloy were used to create the vessels with varying degrees of stenosis. The use of novel inserts significantly reduced the time, materials and cost required in the fabrication of these anatomically realistic phantoms. To prevent residual metal remaining in the final phantom lumens a technique employing clingfilm was used to remove all molten metal from the lumen. These novel flow phantoms developed using efficient methods for producing vessels with various degrees of stenosis can provide a means of evaluation of current and emerging ultrasound technology.