Related Experiment Video
Updated: Jun 11, 2026

A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Aggregate national experience with the wearable cardioverter-defibrillator: event rates, compliance, and survival
Mina K Chung1, Steven J Szymkiewicz, Mingyuan Shao
1Department of Cardiovascular Medicine, Cleveland Clinic, Cleveland, Ohio 44195, USA. chungm@ccf.org
Insights
Patient compliance with the wearable cardioverter-defibrillator (WCD) was satisfactory, with over half using it more than 90% of the time. The WCD demonstrated effective treatment for ventricular tachycardia/ventricular fibrillation (VT/VF), showing survival comparable to implantable cardioverter-defibrillators.
Area of Science:
- Cardiology
- Medical Devices
- Electrophysiology
Background:
- Sudden cardiac death prevention relies on effective antiarrhythmic treatment, patient adherence, and timely intervention for ventricular tachycardia/ventricular fibrillation (VT/VF).
- The wearable cardioverter-defibrillator (WCD) offers a non-invasive option for managing arrhythmia risk.
Purpose of the Study:
- To evaluate patient compliance with the wearable cardioverter-defibrillator (WCD).
- To assess the effectiveness of the WCD as an antiarrhythmic treatment.
- To compare survival outcomes with implantable cardioverter-defibrillator (ICD) patients.
Main Methods:
- A nationwide registry collected data on post-market WCD usage and patient events.
- Survival analysis was performed using the Social Security Death Index.
- WCD patient survival was compared to that of implantable cardioverter-defibrillator (ICD) patients.
Main Results:
- Over 3,500 patients used the WCD, with 52% achieving over 90% daily wear time.
- Eighty sustained VT/VF events occurred, with a 99% first-shock success rate.
- Overall survival during WCD use was 99.2%, with 90% survival for VT/VF events.
Conclusions:
- Satisfactory patient compliance was observed, with significant wear time in over half of patients.
- The WCD demonstrated effective treatment for VT/VF, achieving survival rates comparable to ICDs.
- Asystole emerged as a significant factor in sudden cardiac arrest mortality.
Objectives:
The purpose of this study was to determine patient compliance and effectiveness of antiarrhythmic treatment by the wearable cardioverter-defibrillator (WCD).
Background:
Effectiveness of the WCD for prevention of sudden death is dependent on event type, patient compliance, and appropriate management of ventricular tachycardia/ventricular fibrillation (VT/VF).
Methods:
Compliance and events were recorded in a nationwide registry of post-market release WCDs. Survival, using the Social Security Death Index, was compared with survival in implantable cardioverter-defibrillator (ICD) patients.
Results:
Of 3,569 patients wearing the WCD (age 59.3+/-14.7 years, duration 52.6+/-69.9 days), daily use was 19.9+/-4.7 h (>90% of the day) in 52% of patients. More days of use correlated with higher daily use (p<0.001). Eighty sustained VT/VF events occurred in 59 patients (1.7%). First-shock success was 76 of 76 (100%) for unconscious VT/VF and 79 of 80 (99%) for all VT/VF. Eight patients died after successful conversion of unconscious VT/VF (89.5% survival of VT/VF events). Asystole occurred in 23 (17 died), pulseless electrical activity in 2, and respiratory arrest in 1 (3 died), representing 24.5% of sudden cardiac arrests. During WCD use, 3,541 of 3,569 patients (99.2%) survived overall. Survival occurred in 72 of 80 (90%) VT/VF events and 78 of 106 (73.6%) for all events. Long-term mortality was not significantly different from first ICD implant patients but highest among patients with traditional ICD indications.
Conclusions:
Compliance was satisfactory with 90% wear time in >50% of patients and low sudden death mortality during use. Survival was comparable to that of ICD patients. However, asystole was an important cause of mortality in sudden cardiac arrest events.