Related Experiment Video
Updated: Jun 10, 2026

07:45
The Use of Mixed Reality in Custom-Made Revision Hip Arthroplasty: A First Case Report
Published on: August 4, 2022
A custom-made guide-wire positioning device for hip surface replacement arthroplasty: description and first results
Martijn Raaijmaakers1, Frederik Gelaude, Karla De Smedt
1Department of Reconstructive Hip Surgery, UZ Pellenberg, Katholieke Universiteit Leuven, Belgium. martijn.raaijmaakers@gmail.com
BMC Musculoskeletal Disorders
|July 16, 2010
Summary
A new custom jig accurately positions guide wires for hip surface replacement arthroplasty (SRA), improving femoral component placement. This user-friendly device offers a viable alternative to complex navigation systems.
Area of Science:
- Orthopedic Surgery
- Biomedical Engineering
- Medical Device Design
Background:
- Hip surface replacement arthroplasty (SRA) is an alternative to total hip arthroplasty.
- Optimal femoral component size and position are critical for SRA success.
- Accurate intraoperative placement relies on guide wire positioning, often aided by visual inspection or Computer Assisted Surgery (CAS).
Purpose of the Study:
- To present a custom-made neck jig device for hip surface replacement arthroplasty.
- To evaluate the accuracy and usability of this novel jig in a pilot study.
Main Methods:
- A custom jig was designed based on CT-scan data and surgeon recommendations.
- The jig fits the anterior femoral head and neck, guiding the drill for accurate wire placement.
- The device was tested in 5 patients undergoing total hip arthroplasty, with analysis of guide wire accuracy and surgeon feedback.
Main Results:
- Surgeons rated the jig's stability and fit positively.
- No guide failures occurred, and the device was rated as user-friendly (rated 1-2 for complexity).
- All guide wires were accurately positioned, with maximum angular deviation of 2.9 degrees and 2.1 mm deviation in insertion points.
Conclusions:
- The custom-made jig is a user-friendly and effective tool for SRA.
- It enables accurate guide wire placement, aligning with preoperative plans.
- This device offers a practical solution for precise femoral component positioning in SRA.