The influence of self-absorption on PET and PET/CT shielding requirements
M Elschot1, T C de Wit, H W A M de Jong
1Department of Radiology and Nuclear Medicine, University Medical Center Utrecht, PO. Box 85500, 3508 GA Utrecht, The Netherlands. m.elschot@umcutrecht.nl
Purpose:
The high energy (511 keV) annihilation photons used in positron emission tomography (PET) imaging generally require a substantial amount of lead to protect personnel and the general public from ionizing radiation. A cost-effective design of the PET facility that ensures radiation does not exceed formal dose limits requires accurate estimation of the necessary PET shielding. The American Association of Physicists in Medicine (AAPM) Task Group 108 recently published broad beam transmission factors based on Monte Carlo calculations of 511 keV photons. In this work, an extension to the AAPM model is presented, based on Monte Carlo simulations including the effects of self-absorption on the photon energy spectrum.
Methods:
Monte Carlo calculations were performed using MCNPX. The photon energy spectrum after self-absorption was computed by simulating a normal 18FDG activity distribution in an anthropomorphic phantom. This spectrum was used to calculate the dose rate transmission factors for various wall thicknesses of lead, concrete, and iron. The method was validated by measurement and corresponding simulation of the transmission factors of an 18FDG source in air and in PMMA. Furthermore, a method to generate 3D area dose rate maps of PET facilities incorporating the calculated transmission tables is presented and applied to several shielding situations.
Results:
The calculated self-absorption correction factor and the broad beam transmission factors resulting from Monte Carlo simulations of a monoenergetic point source emitting 511 keV photons were in excellent agreement with the results of the AAPM publication (0.66 vs 0.64 and R2 = 0.999, respectively). However, when all radiation physics, i.e., also the effect of self-absorption on the photon energy spectrum, is included in the Monte Carlo calculations, a substantial reduction in required shielding material was found. For example, including all radiation physics leads to 13.3 mm of lead required to obtain a typical transmission factor of 0.1, instead of 16.0 mm of lead when the AAPM data including only the self-absorption correction factor are used. These findings were confirmed by the experimental measurements. The transmission factors produced in this work can be applied in the same manner as those estimated by AAPM to allow for a cost-effective design of PET and PET/CT facilities without violating radiation safety regulations.
Conclusions:
Taking into account the effect of self-absorption on the photon energy spectrum results in more accurate and cost-effective shielding requirement estimations.
More Related Videos
09:21Human Brown Adipose Tissue Depots Automatically Segmented by Positron Emission Tomography/Computed Tomography and Registered Magnetic Resonance Images
Published on: February 18, 2015
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy (PRRT): 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Related Concept Videos
Biological Effects of Radiation
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...