Related Experiment Video
Updated: Jun 10, 2026

Evaluation of Fluid Overload by Bioelectrical Impedance Vectorial Analysis
Published on: August 17, 2022
An impedance threshold device increases blood pressure in hypotensive patients
Stephen W Smith1, Brent Parquette, David Lindstrom
1Department of Emergency Medicine, Hennepin County Medical Center, Minneapolis, Minnesota 55415, USA. Smith253@umn.edu
Background:
The impedance threshold device (ITD-7) augments the vacuum created in the thorax with each inspiration, thereby enhancing blood flow from the extrathoracic venous systems into the heart.
Objectives:
To the best of our knowledge, the ITD-7 has not previously been investigated in hypotensive patients in the emergency department (ED) or the prehospital setting. The objective of this study was to determine whether the ITD-7 would increase systolic arterial pressures in hypotensive spontaneously breathing patients.
Methods:
The ED study was a prospective, randomized, double-blind, sham control design. Patients with a systolic blood pressure ≤ 95 mm Hg were randomized to breathe for 10 min through an active or sham ITD. The primary endpoint was the change in systolic blood pressure measured non-invasively. The prehospital study was a prospective, non-blinded evaluation of the ITD-7 in hypotensive patients.
Results:
In the ED study, the mean ± standard deviation rise in systolic blood pressure was 12.9 ± 8.5 mm Hg for patients (n = 16) treated with an active ITD-7 vs. 5.9 ± 5.9 mm Hg for patients (n = 18) treated with a sham ITD-7 (p < 0.01). In the prehospital study, the mean systolic blood pressure before the ITD-7 was 79.4 ± 10.2 mm Hg and 107.3 ± 17.6 mm Hg during ITD-7 use (n = 47 patients) (p < 0.01).
Conclusion:
During this clinical evaluation of the ITD-7 for the treatment of hypotensive patients in the ED and in the prehospital setting, use of the device significantly increased systolic blood pressure and was safe and generally well tolerated.
Related Concept Videos
Measurement of Blood Pressure
Alterations in Blood Pressure
Hypertension (High blood pressure)
Hypertension occurs when blood pressure readings consistently exceed the normal range. It is diagnosed when systolic blood pressure (the top number, indicating pressure while the heart beats)...
Hypertension I: Introduction
Antihypertensive Drugs: Action of Diuretics
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Antihypertensive Drugs: Angiotensin-Converting Enzyme Inhibitors