Adaptive statistical iterative reconstruction technique for pulmonary CT: image quality of the cadaveric lung on
Masahiro Yanagawa1, Osamu Honda, Shigeyuki Yoshida
1Department of Radiology, Osaka University Graduate School of Medicine, Suita-city, Osaka, 565-0871, Japan. m-yanagawa@radiol.med.osaka-u.ac.jp
Rationale And Objectives:
To evaluate thin-section computed tomography (CT) images of the lung reconstructed using adaptive statistical iterative reconstruction (ASIR) on standard- and reduced-dose CT.
Materials And Methods:
Eleven cadaveric lungs were scanned by multidetector-row CT with two different tube currents (standard dose, 400 mA; reduced dose, 10 mA). The degree of ASIR was classified into six different levels: 0% (non-ASIR), 20%, 40%, 60%, 80%, and 100% (maximum-ASIR). The ASIR (20%, 60%, and 100%) images were compared with the ASIR (0%) images and assessed visually by three independent observers for image quality using a 7-point scale. The evaluation items included abnormal CT findings, normal lung structures, and subjective visual noise. The median scores assigned by the three observers were analyzed statistically. Quantitative noise was calculated by measuring the standard deviation in a circular region of interest on each selected image of ASIR (0%-100%).
Results:
On standard-dose CT, the overall image quality significantly improved with increasing degree of ASIR (P ≤ .009, Wilcoxon signed-ranks test with Bonferroni correction). As ASIR increased, however, intralobular reticular opacities and peripheral vessels tended to be obscure. On reduced-dose CT, the overall image quality of ASIR (100%) was significantly better than that of ASIR (20%) (P ≤ .009). As ASIR increased, however, intralobular reticular opacities tended to be obscure. Using ASIR significantly reduced subjective and quantitative image noise on both standard- and reduced-dose CT (P < .001, Bonferroni/Dunn's method).
Conclusion:
ASIR improves the image quality by decreasing image noise. Maximum-ASIR may be needed for improving image quality on highly reduced-dose CT. However, excessive ASIR may obscure subtle shadows.
More Related Videos
08:05Lung CT Segmentation to Identify Consolidations and Ground Glass Areas for Quantitative Assesment of SARS-CoV Pneumonia
Published on: December 19, 2020
12:32Image Rendering Techniques in Postmortem Computed Tomography: Evaluation of Biological Health and Profile in Stranded Cetaceans
Published on: September 27, 2020
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...