Related Experiment Video
Updated: Jun 10, 2026

Standardized Model of Ventricular Fibrillation and Advanced Cardiac Life Support in Swine
Published on: January 30, 2020
Airway scope vs Macintosh laryngoscope during chest compressions on a fresh cadaver model
Junho Cho1, Hyun Soo Chung, Sung Pil Chung
1Department of Emergency Medicine, Yonsei University College of Medicine, Seoul 120-752, Republic of Korea.
Purposes:
This study compared the Airway scope (AWS) to the Macintosh laryngoscope (ML) during chest compressions on a fresh cadaver.
Methods:
This was a prospective crossover study. The participants who had experiences with AWS were excluded. The participants intubated with randomly assigned AWS or ML on a fresh cadaver during chest compressions. Primary outcome were as follows: time to intubation, ease of intubation (rated by using the visual analog scale [VAS]), and intubation success rate.
Results:
Twenty-five were enrolled. Median time of intubation was similar between the AWS and ML (AWS, 18.5 seconds vs ML, 18.3 seconds; P = .112). The median VAS of AWS and ML were 3.0 and 2.0, respectively (P = .023). There was no failure of intubation. However, participants replied that the AWS was more difficult to use than the ML.
Conclusion:
Considering the lack of experience with the AWS, AWS could be an alternative intubation device during chest compressions after practices with AWS.
More Related Videos
09:16Open Tracheostomy Gastric Acid Aspiration Murine Model of Acute Lung Injury Results in Maximal Acute Nonlethal Lung Injury
Published on: February 26, 2017
07:45Minimally Invasive Murine Laryngoscopy for Close-Up Imaging of Laryngeal Motion During Breathing and Swallowing
Published on: December 1, 2023
Related Concept Videos
Cardiopulmonary Resuscitation II: ACLS Airway Management
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Assessment of Airway, Skin Color, and Use of Accessory Muscles
Introduction
The initial evaluation of a patient's respiratory system...