Related Experiment Video
Updated: Jun 10, 2026

Effective Analysis of Human Exposure Conditions with Body-worn Dosimeters in the 2.4 GHz Band
Published on: May 2, 2018
[Is it really not possible to use electronic personal dosimeters in clinical exposure situations?]
M Borowski1, B Poppe, H K Looe
1Institut für Röntgendiagnostik und Nuklearmedizin, Klinikum Braunschweig. m.borowski@klinikum-braunschweig.de
Purpose:
Due to significant measuring inaccuracies that can occur under certain conditions, the use of electronic personal dosimeters in statutory measurements in X-ray diagnostics is currently legally restricted. The present study investigates the clinically relevant situations in which measurement errors of more then 20 % can be anticipated.
Materials And Methods:
Four series of experiments were made, comparing the results of the electronic personal dosimeter EPD Mk2.3 to those of reference dosimeters (TLDs and diagnostic dosimeters). On the one hand, personal doses have been determined in the routine operation of controlled areas in various departments. On the other hand, measurements on phantoms have been conducted in extreme but realistic situations under radiation protection. Experiments were conducted in unweakened scattered radiation as well as in unattenuated and attenuated direct radiation.
Results:
The tested electronic personal dosimeter type meets the requirements regarding measurement accuracy for "official" personal dosimeters in all of the examined clinically relevant scattered radiation fields. Only if exposed to radiation directly, an underestimation of the dose can occur and can be greater than 90 %.
Conclusion:
In the range of scattered radiation of diagnostic X-ray equipment, even in pulsed fields, the use of electronic personal dosimeters is reasonable. Considerable measurement errors can only arise in radiation fields that are not realistic under regular conditions and even in connection with most accidents.
More Related Videos
06:21Diffuse Optical Spectroscopy for the Quantitative Assessment of Acute Ionizing Radiation Induced Skin Toxicity Using a Mouse Model
Published on: May 27, 2016
08:23An Automated Microscopic Scoring Method for the γ-H2AX Foci Assay in Human Peripheral Blood Lymphocytes
Published on: December 25, 2021
Related Concept Videos
Biological Effects of Radiation
Electronic Distance Measuring Instruments
Pharmaceutical Poisoning: Potential Scenarios