[Investigation on quality of spirometry in 36 large hospitals in China]
Yi Gao1, Jin-ping Zheng, Jia-ying An
1State Key Laboratory of Respiratory Diseases, Guangzhou Institute of Respiratory Diseases, the First Affiliated Hospital of Guangzhou Medical College, Guangzhou 510120, China.
Objective:
To assess the current status of quality control of spirometry reports in China.
Methods:
Spirometry case reports from January 2008 to October 2008 were collected from pulmonary function testing laboratories in 36 large hospitals (provincial or municipal) in China. The quality analysis was performed according to ATS/ERS standardization for measurement of spirometry. The number of reports that met the criteria for quality control was expressed as percentages.
Results:
A total of 345 spirometry test reports were collected. 82.5% (282/342) met the start-of-test criteria for quality control. 333 reports could be analyzed for free of artifacts, of which 65.8% (219/333) were consistent with the criteria of smooth expiration; the remaining reports (114) failed to meet the criteria, for reasons including cough (29/333, 8.7%), premature glottis closure (8/333, 2.4%), gas leakage (29/333, 8.7%), early termination of expiration (26/333, 7.8%), mouthpiece obstruction (6/333, 1.8%), and incomplete exertion of efforts (49/333, 14.7%). 235 reports were analyzed for end-of-test criteria, with 50.6% (119/235) complying to criteria. 22.6% (78/345) of the reports were tested for more than 3 times, among which 65 reports with the data of each manoeuvre could be analyzed for repeatability. 95.4% (62/65) of the reports met the repeatability criteria, which accounted for 18.0%(62/345)of the total reports collected. The rate of the 2 highest FVC (forced vital capacity) and FEV(1) (forced expiratory volume in one second) of less than 150 ml was 95.4% (62/65) and 100.0% (65/65), respectively. The rate of the 2 highest FVC and FEV(1) of less than 100 ml was 92.3% (60/65) and 87.7% (57/65), respectively. Only 7.2% of the reports (25/345) met all of the 4 criteria listed above.
Conclusions:
The quality of spirometry tests in large (provincial or municipal) hospitals in China needs to be improved in the future.
Related Concept Videos
Pulmonary Function Tests
Pulmonary Function Tests are crucial diagnostic tools for assessing respiratory function, particularly in patients with chronic respiratory disorders. They comprehensively evaluate lung volumes, ventilatory function, breathing mechanics, diffusion, and gas exchange. These tests help diagnose pulmonary diseases and play a significant role in monitoring disease progression, evaluating disability, and assessing response to therapy.
PFTs involve using a spirometer, a...
Chronic Obstructive Pulmonary Disease-IV: Assessement and Diagnostic Studies
Medical History
Respiratory System Abnormal Finding I: Inspection and Percussion
Inspection Findings
During an inspection, several findings may suggest the presence of respiratory distress or disease. Pursed-lip breathing, where exhalation is slowed by...
Assessment of Respiration
Subjective Assessment: Nurses interview the patient to gather information directly during the subjective assessment. It includes questions about the individual's medical history, medications, and symptoms, focusing on past respiratory conditions like asthma or COPD,...