Related Experiment Video
Updated: Jun 10, 2026

Infant Auditory Processing and Event-related Brain Oscillations
Published on: July 1, 2015
Differences in distortion product otoacoustic emission phase recorded from human neonates using two popular probes
Carolina Abdala1, Sumitrajit Dhar
1Division of Communication and Auditory Neuroscience, House Ear Institute, 2100 W Third Street, Los Angeles, California 90057, USA. cabdala@hei.org
Abstract:
DPOAE (2f(1)-f(2)) phase was measured across a 3-octave frequency range from two groups of newborns using ER10B+ and ER10C probe microphones. A marked phase shift was noted in the mid-to-high frequency range for newborn data recorded with the ER10C only. In contrast, the ER10B+ produced phase that was approximately invariant as a function of frequency for most of the range. Probe-related phase shifts can be effectively eliminated by correcting for variations in the phases of the primary tones. Results highlight the importance of detecting and correcting for system-related phase shifts so they are not misinterpreted as cochlear in origin.