Invasive validation of a new oscillometric device (Arteriograph) for measuring augmentation index, central blood

Iván G Horváth1, Adám Németh, Zsófia Lenkey

  • 1aHeart Institute, Medical School, University of Pécs, Pécs, Hungary bDepartment of Cardiology, University of Rome La Sapienza, Polo Pontino, Italy.

Insights

A new oscillometric device accurately measures aortic pulse wave velocity (PWVao), aortic augmentation index (Aix), and central systolic blood pressure (SBPao) noninvasively. These findings support its clinical use for assessing cardiovascular health.

Area of Science:

  • Cardiovascular Physiology
  • Biomedical Engineering
  • Diagnostic Technology

Background:

  • Accurate measurement of aortic pulse wave velocity (PWVao), aortic augmentation index (Aix), and central systolic blood pressure (SBPao) is clinically significant.
  • Current methods for obtaining these aortic parameters are often invasive and difficult to access.

Purpose of the Study:

  • To evaluate the accuracy of a new, easily applicable oscillometric device, the Arteriograph.
  • To compare simultaneous noninvasive measurements from the Arteriograph against invasive reference standards.

Main Methods:

  • Invasive measurements of aortic Aix, SBPao, and PWVao were performed during cardiac catheterization.
  • These invasive values were compared with simultaneous measurements obtained using the Arteriograph device.

Main Results:

  • Strong correlations (r=0.9-0.95, P<0.001) were observed between invasive and Arteriograph-derived Aix and SBPao.
  • PWVao measurements showed excellent agreement (Pearson's r=0.91, P<0.001) between invasive and oscillometric methods.
  • Limits of agreement were acceptably low (11.4% for Aix, 1.59 m/s for PWVao) for clinical application.

Conclusions:

  • The Arteriograph device accurately measures aortic Aix, SBPao, and PWVao noninvasively.
  • The device demonstrates strong correlation and low limits of agreement compared to invasive methods.
  • These findings support the clinical utility of the Arteriograph for assessing central hemodynamic parameters.
Abstract

Related Concept Videos

Assessing Blood pressure using a doppler ultrasound01:19

Assessing Blood pressure using a doppler ultrasound

To obtain accurate blood pressure measurements in clinical settings, especially when traditional methods are insufficient, healthcare professionals utilize the Doppler ultrasound technique. This method uses high-frequency sound waves to detect blood flow within the arteries, which is crucial for patients with conditions that complicate circulatory system assessment.
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of blood pressure in brachial artery(two-step method)01:23

Assessment of blood pressure in brachial artery(two-step method)

Measuring blood pressure is a fundamental skill in healthcare that aids in diagnosing and monitoring hypertension and other cardiovascular conditions. An aneroid sphygmomanometer, commonly used in clinical settings, offers a manual and precise method for blood pressure measurement. The technique for using this instrument involves specific steps that must be carefully executed to ensure accuracy. The following detailed description outlines a two-step technique for assessing blood pressure using...
Sites for measuring blood pressure01:21

Sites for measuring blood pressure

Blood pressure measurement is a fundamental clinical procedure, providing crucial data for assessing cardiovascular health. Among the various sites for this measurement, the brachial and popliteal arteries are predominantly utilized due to their accessibility and the reliability of their readings. This lesson delves into the anatomical significance, methodology, and considerations of measuring blood pressure at these locations.
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(one-step method)01:15

Assessment of blood pressure in brachial artery(one-step method)

This procedural guide systematically measures blood pressure using an oscillometric digital sphygmomanometer, emphasizing accuracy, patient safety, and comfort.
Prepare for the Procedure:
Equipments Used To Measure Blood Pressure01:30

Equipments Used To Measure Blood Pressure

Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...