Impact of pulse forming network and injection current parameters on output and energy variations of helical
A Gutierrez1, S Stathakis, C Esquivel
1Department of Radiation Oncology, University of Texas Health Science Center at San Antonio, San Antonio, TX, 78229, USA.
Purpose:
In this study, an experiment was devised to establish the dependency of the impact of pulse forming network (PFN) and injection current (IC) parameters on output and energy variations of helical tomotherapy (HT) on the radiation beam output and energy.
Methods:
HT has unique radiation beam characteristics due to the absence of a flattening filter. As with conventional linear accelerators, the machine output and energy should be within a + or - 2% tolerance according to published studies. However, because a dose servo is not utilized in a HT unit, these parameters may drift out of the + or - 2% tolerance due to various reasons such as high machine temperatures. With this in mind, physicists and engineers must adjust certain machine parameters to reset the output and energy to within the tolerance of the commissioned baseline. Two parameters commonly adjusted are: PFN voltage (V(PFN)) and IC voltage (V(IC)).
Results:
Results showed that the HT unit possesses different working zones defined by the V(PFN) and V(IC) parameter settings. The working zones were classified into 5 zones: 1) low dose rate zone - radiation dose rate much lower than nominal dose rate and machine cannot run due to low dose rate fault; 2) normal dose rate zone - dose rate is within tolerance of nominal dose rate and machine can run without dose rate fault; 3) dose rate failure during treatment zone - dose rate within the tolerance of the nominal dose rate however machine interrupts during treatment due to dose rate fault; 4) high dose rate zone - dose rate is higher than nominal dose rate and machine cannot run due to high dose rate fault; and 5) inoperable dose rate zone - dose rate is much higher than the nominal dose rate and machine cannot run.
Conclusion:
The results of this study may provide a quick guide for physicists to adjust their HT unit V(PFN) and V(IC) values in order to reset the radiation beam output and energy back to within the tolerance of the commissioned baseline.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences
Induced Electric Fields: Applications
Torque On A Current Loop In A Magnetic Field
Consider a rectangular current-carrying loop containing N turns of wire, placed in a uniform magnetic field. The net force on a current-carrying loop...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Imaging Studies III: Computed Tomography