An automated turbulent flow liquid chromatography-isotope dilution mass spectrometry (LC-IDMS) method for
Robert Harlan1, William Clarke, Joseph M Di Bussolo
1John Hopkins Hospital, Baltimore, MD, USA.
Background:
When creatinine concentrations determined by routine clinical assays are in question, reference methods can aid investigation. Currently available reference methods are significantly labor-intensive, which prevents implementation in a routine clinical laboratory.
Methods:
Creatinine D-3 internal standard was added to serum prior to chromatographic separation. A TurboFlow Cyclone MCX column was used for online solid phase extraction (SPE) to remove large biomolecules such as proteins, carbohydrates and phospholipids from the serum specimen. Creatinine and creatinine D-3 were then eluted onto a Hypercarb (porous graphitic carbon) column for separation. Analytes were detected using electrospray ionization tandem mass spectrometry and measured by monitoring parent ions of m/z 114 and 117, respectively.
Results:
Total precision at multiple levels was found to be less than 6% (1.0-7.5 mg/dl). Limit of detection (LOD) was 0.05 mg/dl and limit of quantitation (LOQ) was 0.20 mg/dl. Average recovery was 107.5% (0.37-5.95 mg/dl). Analysis of standard reference materials from The National Institute of Standards and Technology (NIST) confirmed accuracy of the method. No significant difference was found between the liquid chromatography-isotope dilution mass spectrometry (LC-IDMS) method and the Roche Creatinine Plus enzymatic assay.
Conclusion:
The automated turbulent flow LC-IDMS method for quantitation of serum creatinine is accurate, robust, and easy to perform and may serve as a quick and inexpensive alternative to current creatinine reference methods.
More Related Videos
07:34Large Scale Non-targeted Metabolomic Profiling of Serum by Ultra Performance Liquid Chromatography-Mass Spectrometry (UPLC-MS)
Published on: March 14, 2013
09:26Cellular Lipid Extraction for Targeted Stable Isotope Dilution Liquid Chromatography-Mass Spectrometry Analysis
Published on: November 17, 2011
Related Concept Videos
Drug Dosing in Renal Diseases: Measurement of Serum Creatinine Concentration and Clearance
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview