Objectives:
This study was designed to assess the repeatability of the Emed® ST2 system and identify the range of pressure values observed in the normal foot.
Methods:
Measurements were taken from twenty-three healthy subjects, 14 females and 9 males, on two occasions 7 days apart. Begin of contact (BC), end of contact (EC), contact time (CT), peak pressure (PP), instant of peak pressure (IPP), contact area (CA) and pressure-time integral (PTI) were recorded.
Results:
The coefficient of repeatability (CR) was less than 16.0% for all 63 parameters considered. In 87.3% of the parameters investigated (55 of 63) the CR (expressed as a percentage of the mean) was less than 10%. The highest areas of PP were found under the great toe and second metatarsal heads, with mean (S.D.) equal to 435 kPa (202) and 407 kPa (146), respectively, followed by the third metatarsal head 345 kPa (96) and the hindfoot 332 kPa (93). The CT (% ROP (range of pressure)) was in the range 74-85% under the metatarsal heads, and 71% under the great toe. CA was highest under the heel at 33.8 cm(2).
Conclusion:
Emed® ST2 system was found to be repeatable. The presented range of parameters compared very well to the results presented in the literature for the Emed® ST4 system.
Related Concept Videos
Assessment of blood pressure in brachial artery(two-step method)
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
Errors occurring during blood pressure monitoring
Several factors...
Measurement of Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.