Ventricular oversensing due to manufacturer-related differences in implantable cardioverter-defibrillator signal
Joachim Seegers1, Markus Zabel, Lars Lüthje
1Department for Cardiology, University Medical Center Goettingen, Germany.
Aims:
Ventricular oversensing remains a significant problem in modern implantable cardioverter-defibrillator (ICD) systems. The role of manufacturer-related differences in device-based signal processing on the one hand and sensing lead properties on the other is largely unknown. This retrospective study aimed to evaluate and compare the incidence and mechanisms of ventricular oversensing in single- or dual-chamber ICD systems from Biotronik (BTK), Guidant (GDT), and Medtronic (MDT).
Methods And Results:
In 245 consecutive patients, device function, stored episodes, and lead parameters were evaluated at implant and during 243 ± 18 days of follow-up. Oversensing occurred in 3.3% of the patients [four T-wave oversensing (TWO), three R-wave double-sensing, and one diaphragmatic myopotentials]. The incidence of ventricular oversensing was generally low if an ICD device and an ICD lead were from one identical manufacturer (BTK: 2/68, GDT: 1/80, and MDT: 1/69; P = 0.67). In the subgroup of ICDs connected to leads from other manufacturers, however, ventricular oversensing was observed in none of 19 GDT and MDT ICDs but in 4 of 9 BTK devices (44.4%, P = 0.012). These four cases of oversensing (three R-wave double-sensing and one TWO) all occurred in BTK ICDs connected to integrated bipolar ICD leads from other manufacturers.
Conclusion:
Significant manufacturer-related differences exist in the incidence of ventricular oversensing in single- and dual-chamber ICD systems. Biotronik devices connected to integrated bipolar ICD leads from other manufacturers have an increased risk for R-wave double-sensing.
Related Concept Videos
Mechanism of Cardiac Arrhythmias
Imbalances in Cardiac Output
CHF can occur due to the failure of either side of the heart. Left-side failure leads to pulmonary congestion—the right side continues to send blood...
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin to...