An Improved Mechanical Testing Method to Assess Bone-implant Anchorage
Published on: February 10, 2014
Monitoring of fixture osteointegration after BAHA® implantation
Fabrizio Clemente1, Michele Costa, Simonetta Monini
1Department of Neuroscience, Mental Health and Sensory Organs, II Medical School Sapienza University Rome, Italy.
Abstract:
The osteointegration phase of BAHA(®) fixture was assessed via electrical impedance spectroscopy in BAHA(®) implantees. Measurements were carried out by using a prototype device and were correlated with previously published histological data obtained at certain times after surgery, i.e. 1-2 days, 1 week, 1 and 3 months. Module variations of impedance spectra were found to be in agreement with the different pathophysiological conditions of bone ingrowth. The proposed methodology has shown to be promisingly reliable to properly monitor BAHA fixture osteointegration.