Anterior segment optical coherence tomography measurement of flap thickness after myopic LASIK using the Moria one
Hui-Jin Chen1, Ying-Jie Xia, Yan-Ying Zhong
1Peking University Third Hospital, Peking University Eye Center, Beijing, China.
Purpose:
To analyze the accuracy and consistency of corneal flap thickness created in myopic LASIK using the Moria One Use-Plus microkeratome compared with the Moria M2 Single Use 90-microm microkeratome.
Methods:
Bilateral LASIK was performed in 68 myopic patients. Flaps were created using the One Use-Plus microkeratome in 82 eyes (41 patients) and the M2 90-microm microkeratome in 54 eyes (27 patients). Flap complications and visual outcomes were evaluated. Horizontal "High Res. Corneal" scan pattern of anterior segment optical coherence tomography (AS-OCT) was applied to measure flap thickness at five locations (0, +/-2, and +/-3.5 mm from the corneal vertex) on the first postoperative day.
Results:
No significant differences were noted in flap complications and visual outcomes between groups. The central flap thickness was dramatically thinner in the One Use-Plus group (114.7+/-10.1 microm and 109.4+/-11.0 microm for right and left eyes, respectively) than in the M2 group (155.6+/-14.8 microm and 151.6+/-12.5 microm for right and left eyes, respectively) (P<.001). The One Use-Plus did not show a markedly better uniformity than the M2; the variation was mainly observed in the periphery. Multiple linear regression showed that for the One Use-Plus, the steeper the preoperative keratometry, the thicker the flap thickness, and for the M2, the thicker the preoperative pachymetry, the thicker the flap (P<.1).
Conclusions:
The One Use-Plus and M2 microkeratomes have similar safety and efficacy. The flap created by the One Use-Plus was much thinner than the flap created with the M2; however, the One Use-Plus can not realize a fully planar-shaped flap.