Published on: October 17, 2013
A nonradiographic system for assessing pressure for the Codman-Hakim Programmable Valve
Nancy Olsen Bailey1, Mark Luciano, Michael V Ward
1Department of Neurosurgery, Brigham and Women's Hospital, Boston, Massachusetts 02115, USA.
A new system accurately verifies Codman-Hakim Programmable Valve settings without radiation. This method is efficient, but cranial radiographs may still be needed for about one-third of patients.
Area of Science:
- Neurosurgery
- Medical Device Technology
- Clinical Engineering
Background:
- The Codman-Hakim Programmable Valve is a common device for managing hydrocephalus.
- Accurate valve setting is crucial for effective treatment.
Purpose of the Study:
- To develop and test a non-radiographic system for verifying Codman-Hakim Programmable Valve settings.
- To improve the efficiency and safety of valve adjustment procedures.
Main Methods:
- A prospective clinical trial was conducted across nine research sites.
- A novel Valve Programming and Verification System was evaluated.
Main Results:
- The new system successfully established the valve setting in 62% of subjects with a programmable valve.
- The system offers a radiation-free alternative for valve verification.
Conclusions:
- The Valve Programming and Verification System is cost-effective and time-efficient.
- While beneficial, cranial radiographs remain necessary for approximately one-third of patients requiring adjustment completion.
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Pressure Gauges
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...