Related Experiment Video
Updated: Jun 10, 2026

Endoscopic Cholesteatoma Surgery
Published on: January 19, 2022
Optical coherence tomography of cholesteatoma
Hamid R Djalilian1, Marc Rubinstein, Edward C Wu
1Department of Otolaryngology-Head and Neck Surgery, University of California Irvine, Irvine, California 92868, USA. hdjalili@uci.edu
Objectives:
To image cholesteatoma using optical coherence tomography (OCT) and correlate the results with clinical findings and conventional observations obtained using binocular microscopy and histology. OCT is a high-resolution optical imaging modality that generates cross-sectional images of turbid media, such as tissue with resolution approaching that of light microscopy. OCT relies on intrinsic differences in tissue optical properties for image contrast.
Study Design:
In vivo prospective clinical study.
Setting:
University Medical Center.
Patients:
Patients with cholesteatoma undergoing otologic surgery.
Intervention:
Using a commercial OCT imaging system, we obtained cross-sectional images (resolution, approximately 10 microm; depth penetration, approximately 1 mm) of cholesteatomas.
Main Outcome Measures:
Images are obtained by raster scanning a single mode fiber across the interior of the probe. The imaging probe is sterilized and inserted into the middle ear or mastoid under microscopic guidance, and still images of the middle ear or mastoid mucosa and cholesteatoma when present were obtained.
Results:
OCT images of cholesteatomas demonstrate differences in signal intensity, which are distinct from those of normal or inflamed middle ear/mastoid mucosa. Identification of keratin in cholesteatoma, even if very thin, distinguished it from inflamed mucosa.
Conclusion:
This is the first study that systematically used OCT to image cholesteatoma during otologic surgery. Cholesteatomas can be distinguished from normal or inflamed adjacent mucosa.