Atrial pacing or ventricular backup-only pacing in implantable cardioverter-defibrillator patients
Michael O Sweeney1, Kenneth A Ellenbogen, Anthony S L Tang
1Cardiac Pacing and Heart Failure Device Therapies, Brigham and Women's Hospital, Boston, Massachusetts 02115, USA. mosweeney@partners.org
Background:
The need for pacing support in typical ICD patients is unknown.
Objective:
This study sought to determine whether atrial pacing with ventricular backup pacing is equivalent to ventricular backup pacing only in implantable cardioverter-defibrillator (ICD) patients.
Methods:
We randomized 1,030 patients from 84 sites with indications for ICDs, with sinus rhythm, and without symptomatic bradycardia to atrial pacing with ventricular backup at 60 beats/min (518) or ventricular backup pacing at 40 beats/min (512). The primary end points were time to death, heart failure hospitalization (HFH), and heart failure-related urgent care (HFUC).
Results:
Follow-up was 2.4 ± 0.8 years when the trial was stopped for futility. There were 355 end point events (103 deaths, 252 HFH/HFUC) in 194 patients favoring ventricular backup pacing (event-free rate 77.7% vs. 80.3% for atrial pacing at 30 months; hazard ratio 1.14, upper confidence bound 1.59, prespecified noninferiority threshold 1.21), therefore equivalence between pacing arms was not demonstrated. Overall HFH/HFUC rates were slightly higher during atrial pacing (event-free rate 85.4% vs. 86.4% for ventricular backup pacing). Exploratory analyses revealed that the difference in HFH/HFUC rates was largely seen in patients with a PR interval ≥230 ms. There were no differences between groups for atrial fibrillation, ventricular tachycardia/ventricular fibrillation, quality of life, or echocardiographic measurements. Fewer patients in the atrial pacing group were reported to develop an indication for bradycardia pacing (3.7% vs. 7.3%, P = .0053).
Conclusion:
Equivalence between atrial pacing and ventricular backup pacing only could not be demonstrated.
Clinical Trials Identifier:
NCT00281099.
Related Concept Videos
Cardiomyopathy V: Interprofessional Care
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Cardiopulmonary Resuscitation III: AED Use
Dysrhythmias VI: Management of Dysrhythmias
Cardiomyopathy II: Dilated Cardiomyopathy