Related Experiment Video
Updated: Jun 10, 2026

Investigating the Three-dimensional Flow Separation Induced by a Model Vocal Fold Polyp
Published on: February 3, 2014
Gas distribution in a two-compartment model ventilated in high-frequency percussive and pressure-controlled modes
Umberto Lucangelo1, Agostino Accardo, Alessandro Bernardi
1Department of Perioperative Medicine, Intensive Care and Emergency, Cattinara Hospital, Trieste University School of Medicine, Strada di Fiume 447, 34139, Trieste, Italy. u.lucangelo@fmc.units.it
Purpose:
To demonstrate in a two-compartment heterogeneous mechanical model of the lung how different loads applied to one compartment, while the other is kept constant, would modify gas distribution between the two pathways under high-frequency percussive ventilation (HFPV). Additionally, these results were compared with those generated in the same model by pressure-controlled ventilation (PCV).
Methods:
Analysis was based on a Siemens lung simulator, representing a fixed branch of the system with an elastance equal to 45 cmH(2)O/L and a resistance of 20 cmH(2)O/L/s, and a single-compartment lung simulator, representing a variable pathway of the model, presenting three elastic loads varying between 35 and 85 cmH(2)O/L and three resistive loads varying between 5 and 50 cmH(2)O/L/s. Each simulator represented one compartment of the model connected to a central airway that was ventilated with either a volumetric diffusive respirator (VDR-4; Percussionaire Corporation, Sandpoint, ID, USA) or a Siemens Servo 900c ventilator. Flow and pressures were measured in each branch of the model under nine conditions representing the combinations of three elastic and three resistive loads (variable branch) while the loads in the other pathway were kept constant.
Results:
HFPV was able to avoid hyperinflation and reduce tidal volume in a bicompartmental heterogeneous lung model. Under HFPV, gas distribution between the two compartments was not constrained by their time constants. PCV yielded gas distribution as determined by the time constant of each compartment.
Conclusions:
HFPV accommodated volume distribution without overinflating compartments with low time constants, thus possibly presenting a potential protective behavior in mechanically heterogeneous lungs.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation I: Indication and Settings
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Compartment Models: Two-Compartment Model
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)