Successful use of the five-in-six system with IVUS guided transradial renal artery stenting
1Department of Cardiology, Cardiology Department, National Heart Institute, Cairo, Egypt.
Insights
The Terumo "five-in-six" system offers a solution for transradial renal interventions by extending guiding catheter (GC) length. This technique provides crucial backup support for renal stenting, especially in challenging anatomies.
Area of Science:
- Interventional Cardiology
- Vascular Surgery
- Medical Device Innovation
Background:
- Transradial renal interventions face challenges due to the long distance between vascular access and target lesions.
- Standard guiding catheters (GCs) may lack sufficient length for engaging renal arteries in certain patients.
Observation:
- A novel approach using the Terumo "five-in-six" system was employed to extend the working length of a 5 Fr GC within a 6 Fr GC.
- This system provided the necessary extra length and enhanced backup support for renal artery engagement.
Findings:
- The "five-in-six" system successfully facilitated renal stenting under intravascular ultrasound (IVUS) guidance.
- The super soft distal segment of the 5 Fr GC allowed deep intubation into the renal artery, ensuring stability.
Implications:
- The "five-in-six" system presents a viable alternative strategy for transradial renal artery interventions requiring extended reach or improved backup support.
- This technique can enhance procedural success rates in complex renal interventions.
Abstract:
In transradial renal intervention, the long distance between the vascular access and the target lesion represents a major issue. The 100 cm guiding catheter (GC) may be too short to engage the renal artery particularly in tall patients or with excessive aortic turtousity. On the other hand, with the use of 125 cm long GC, only few premounted dedicated renal stents can be used. The shaft of the stent delivery system must be 140 cm or more to be compatible with the 125 GC. However, in many renal interventions, only few more centimeters are required to give the GC the sufficient working length needed for proper engagement and back up support. For this purpose, we used the Terumo "five-in-six" system to provide the necessary extra length through the protruding part of the 5 Fr GC out of the tip of the 6 Fr GC. With the super soft distal segment of the 5 FR GC deeply intubated into the right renal artery, we had the sufficient backup support to do renal stenting with IVUS guidance. This case illustrates that the "five-in-six" system can be an alternative strategy whenever extra length or strong backup support is needed during transradial renal artery intervention.