Reproducibility of intraocular lens decentration and tilt measurement using a clinical Purkinje meter
Yutaro Nishi1, Nino Hirnschall, Alja Crnej
1Department of Pathology, Institute of Ophthalmology, University College London, and Moorfields Eye Hospital NHS Foundation Trust, London, United Kingdom.
Purpose:
To determine the reproducibility of intraocular lens (IOL) decentration and tilt measurements with a new Purkinje meter instrument.
Setting:
Moorfields Eye Hospital NHS Foundation Trust, London, United Kingdom.
Methods:
After pupil dilation, images of pseudophakic eyes with a plate-style IOL (Akreos Adapt) were obtained using a recently developed Purkinje meter. Intraocular lens decentration and tilt were evaluated by analyzing the captured images using a semiobjective method by marking the reflexes in the images and automatic calculation using a dedicated software program. In study 1, examiner 1 examined the eyes first followed by examiner 2. Ten minutes later, examiner 1 performed a second measurement, after which the intraexaminer and interexaminer reproducibility were determined. In study 2, a Purkinje meter was used to measure pseudophakic eyes with slitlamp finding of clinical IOL decentration, IOL tilt, or both. The results were compared with retroillumination photographs and slitlamp findings.
Results:
In study 1, there was high intraexaminer reproducibility for decentration (r = 0.95) and tilt (r = 0.85) and high interexaminer reproducibility for decentration (r = 0.84) and tilt (r = 0.75). In study 2, even in extreme cases of decentration and/or tilt, the Purkinje meter measurements were possible and appeared to correlate well with slitlamp findings.
Conclusions:
Acquisition of images in pseudophakic eyes with the Purkinje meter was simple and rapid. The method was highly reliable for 1 examiner and between 2 examiners.