[Usefulness of the collimator detector response (CDR) recovery and the scatter correction by the effective scatter
Yuuya Nakamura1, Seiji Tomiguchi, Noboru Katsuda
1Kumamoto University, Japan.
Unlabelled:
The aim of this study was to evaluate the usefulness of the collimator detector response (CDR) recovery and the effective scatter source estimation (ESSE) method which is the scatter correction method built into the ordered subsets expectation maximization (OSEM) method.
Method:
The SPECT quality evaluation phantom and the anthropomorphic torso phantom were used in this study, and image contrast and uniformity were evaluated. The effect of each image correction method on the quantification of absolute radioactivity was also assessed.
Results And Conclusion:
Image contrast and uniformity were improved with the combination of the CDR recovery and triple energy window (TEW) method or the ESSE method. The combination of the CDR recovery and the ESSE method was the best method for the estimation of absolute radioactivity. Image quality of the SPECT is improved by the combination of CDR recovery and scatter correction in addition to attenuation correction. CDR recovery in addition to attenuation and scatter correction is also useful for the quantification of absolute radioactivity.
More Related Videos
11:09High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
11:13Quantification of Mouse Heart Left Ventricular Function, Myocardial Strain, and Hemodynamic Forces by Cardiovascular Magnetic Resonance Imaging
Published on: May 24, 2021
Related Concept Videos
Acute Coronary Syndrome III: Diagnostic Studies
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies for Cardiovascular System V: CT