Relationship between measurement site and motion artifacts in wearable reflected photoplethysmography
Yuka Maeda1, Masaki Sekine, Toshiyo Tamura
1Department of Medical System Engineering, Chiba University, Chiba, Japan. y.maeda@graduate.chiba-u.jp
Abstract:
Pulse rates obtained from wearable photoplethysmography (PPG) sensors are important for monitoring cardiovascular condition, especially during exercise. However, it is difficult to precisely count pulse rates during exercise because PPG is sensitive to body movement. The artifacts from body movement are caused by a change in the blood volume at the measurement site, in addition to pulsatile changes. Here, we investigated the influence of motion artifact with respect to light source and anatomical sites. In this study, we compared the signal from green-light PPG to that from infrared PPG at different anatomical sites. In these experiments, 11 subjects were asked to either assume a resting position or generate spontaneous motion artifact by jumping and swinging their arm. As a result, pulse rates obtained from green-light PPG showed a higher correlation with the ECG R-R interval as compared to those obtained with infrared. Additionally, the signal from the upper arm showed less artifact than did the peripheral one. Therefore, the green-light PPG may be useful for pulse rate monitoring.
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...