[Coma and resolution in wide spectral region Czerny-Turner spectrometer]
Tan-Xuan Chen1, Huai-Dong Yang, Ke-Xin Chen
1State Key Laboratory of Precision Measurement Technology and Instruments, Tsinghua University, Beijing 100084, China. cth03@mails.thu.edu.cn
Abstract:
The Czerny-Turner layout, which is most frequently used in miniature spectrometers, should follow Shafer's coma-free condition and Fastie's flat-field principal to eliminate the central wave's primary coma and maximize its resolution. However, the design process does not take the comas and resolutions at non-central waves into consideration. Based on the theory of primary coma in reflection optical system, the present paper points out that in the crossed beam design, the resolutions at wide 'spectral region present a "V" shape, while in the M design, the resolutions change little over the whole region, and present an approximately straight line shape, so the latter kind of spectrometer maintains a far more consistent resolution than the former one. Accordingly, this paper designs two kinds of spectrometers with spectrum regions from 400 to 600 nm, and carries out theoretical simulation and contrast experiment. The result demonstrates that for the two designs the resolutions at the fringe wavelength are 3.8 times and 1.5 times respectively that at the central wavelength, which accords with the conclusion of the theoretical simulation.
Related Concept Videos
Atomic Emission Spectroscopy: Interference
UV–Vis Spectroscopy: Molecular Electronic Transitions
Atomic Emission Spectroscopy: Instrumentation
IR Spectroscopy: Molecular Vibration Overview
Stretching vibrations are vibrational motions that occur along the bond line, changing the bond length or distance between two bonded atoms. They are further distinguished as symmetric or asymmetric. In symmetric stretching, the...
Atomic Emission Spectroscopy: Overview
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...