Related Experiment Video
Updated: Jun 10, 2026

Minimal Invasive Resection of Large Retrosternal Thyroid Goiter
Published on: September 20, 2024
Management of tracheal obstruction caused by benign or malignant thyroid disease using covered retrievable
Won-Kyeong Kim1, Ji Hoon Shin, Jin Hyoung Kim
1Department of Internal Medicine, Chuncheon Armed Forces Hospital, University of Ulsan College of Medicine, Asan Medical Center, Seoul, Korea.
Background:
Curative resection is often contraindicated in patients with airway obstruction by aggressive thyroid disease.
Purpose:
To evaluate the safety and clinical effectiveness of covered retrievable self-expandable nitinol stents placed in patients with airway obstruction caused by benign or malignant thyroid disease.
Material And Methods:
From 1996 to 2009, covered retrievable self-expandable nitinol stents were placed in nine symptomatic patients with malignant (n=7) or benign (n=2) thyroid disease. Improvement in patient respiratory status, complications and their management, and survival data were evaluated. Stents were removed if stent-related complications occurred or the stents were no longer necessary.
Results:
A total of 11 stents were successfully placed without procedure-related complications. Improvement of more than one dyspnea grade was seen in eight of nine patients (89%), while the ninth patient underwent intubation due to upper airway swelling. Stent migration, sputum retention, and tumor overgrowth occurred in 27.3% (n=3), 9.1% (n=1), and 9.1% (n=1) of the stents, respectively. A total of six stents were successfully removed because of stent-related complications (n=4) or after total thyroidectomy (n=2). In two patients with benign thyroid goiters, a stable airway was maintained until total thyroidectomy.
Conclusion:
Placement of covered retrievable self-expandable nitinol stents was safe and effective in patients with airway obstruction caused by benign or malignant thyroid disease. Stent retrievability was very useful in cases of stent-related complications. In patients with benign thyroid disease, stent placement can serve as an effective bridge to surgery.
Related Concept Videos
Trachea
Anatomical Features:
Location: About half of the trachea is situated in the neck, anterior to the esophagus, and extends from the larynx (at the level of the...
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Tracheostomy Care I: Pre-procedural Steps
Required Equipment
The equipment necessary for tracheostomy care includes: