Agreement between directly measured blood pressure and pressures obtained with three veterinary-specific
Mark J Acierno1, Diana Seaton, Mark A Mitchell
1Department of Veterinary Clinical Science, School of Veterinary Medicine, Louisiana State University, Baton Rouge, LA 70803, USA. macierno@vetmed.lsu.edu
Objective:
To determine whether veterinary-specific oscillometric blood pressure units yield measurements that are in good agreement with directly measured blood pressures in cats.
Design:
Evaluation study.
Animals:
21 cats undergoing routine spaying or neutering.
Procedures:
A 24-gauge catheter was inserted in a dorsal pedal artery, and systolic, diastolic, and mean arterial pressures were directly measured with a validated pressure measurement system. Values were compared with indirect blood pressure measurements obtained with 3 veterinary-specific oscillometric blood pressure units.
Results:
There was poor agreement between indirectly and directly measured blood pressures. For unit 1, bias between indirectly and directly measured values was -14.9 mm Hg (95% limits of agreement [LOA], -52.2 to 22.4 mm Hg), 4.4 mm Hg (95% LOA, -26.0 to 34.8 mm Hg), and -1.3 mm Hg (95% LOA, -26.7 to 24.1 mm Hg) for systolic, diastolic, and mean arterial pressures, respectively. For unit 2, bias was -10.3 mm Hg (95% LOA, -52.9 to 32.2 mm Hg), 13.0 mm Hg (95% LOA, -32.1 to 58.0 mm Hg), and 9.1 mm Hg (95% LOA, -32.9 to 51.2 mm Hg) for systolic, diastolic, and mean arterial pressures, respectively. For unit 3, bias was -13.4 mm Hg (95% LOA, -51.8 to 25.1 mm Hg), 8.0 mm Hg (95% LOA, -25.5 to 41.6 mm Hg), and -3.6 mm Hg (95% LOA, -31.6 to 24.5 mm Hg) for systolic, diastolic, and mean arterial pressures, respectively.
Conclusions And Clinical Relevance:
Results suggested that none of the 3 veterinary-specific oscillometric blood pressure units could be recommended for indirect measurement of blood pressure in cats.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Pre-Procedural Guidelines for Assessing Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)