A prospective randomized trial comparing supraglottic airways for flexible bronchoscopy in children
Related Experiment Video
Updated: Jun 10, 2026

The Flexible Rhino-Laryngoscope for Awake Nasotracheal Intubation
Published on: August 2, 2024
Paul A Baker1, Katharine E J Brunette, Catherine A Byrnes
1Department of Paediatric Anaesthesia, Starship Children's Health, Auckland, New Zealand. pbaker@adhb.govt.nz
Insights
For pediatric flexible bronchoscopy, single-use polyvinyl chloride supraglottic airways (SGA) showed higher resistance than reusable silicone SGAs. The silicone LMA Classic and Boss Systems masks were most effective for bronchoscope manipulation.
Area of Science:
- Pediatric Anesthesiology
- Respiratory Medicine
- Medical Device Engineering
Background:
- Supraglottic airways (SGAs) are standard for pediatric flexible bronchoscopy, serving as ventilation and conduit devices.
- Some disposable SGAs present higher resistance to bronchoscope manipulation than reusable LMA Classic (cLMA).
Purpose of the Study:
- To compare the performance of different SGAs during pediatric flexible bronchoscopy.
- To evaluate bronchoscope manipulation resistance, insertion time, and visualization across various SGAs.
Main Methods:
- A prospective randomized controlled trial involving 100 children.
- Comparison of cLMA, Ambu Aura Once, Portex Soft Seal, Boss Systems, and LMA Unique.
- Assessment of subjective resistance (linear analog scale), insertion time, laryngeal view, and cuff pressures.
Main Results:
- Polyvinyl chloride (PVC) disposable SGAs (Ambu, Unique, Portex) exhibited significantly higher resistance to bronchoscope manipulation than the silicone cLMA (P < 0.0001).
- Ambu and Unique SGAs were clinically inferior to cLMA across all airway levels.
- Portex Soft Seal showed inferiority below the larynx (P < 0.006) and overall (P < 0.007).
- Boss Systems single-use laryngeal mask demonstrated comparable effectiveness to the cLMA.
Conclusions:
- Single-use PVC laryngeal masks are inferior to the silicone cLMA and Boss Systems masks for pediatric flexible bronchoscopy.
- Silicone-based SGAs, like cLMA and Boss Systems, offer superior performance in terms of reduced resistance and better manipulation during the procedure.
Objective:
A prospective randomized, controlled trial was conducted comparing supraglottic airways (SGA) for flexible bronchoscopy in 100 children.
Background:
Pediatric flexible bronchoscopy is commonly performed using a SGA as both a ventilation device and a conduit for flexible bronchoscopy. We observed that some disposable SGAs were associated with increased resistance to bronchoscope manipulation compared to the LMA Classic (cLMA).
Methods:
We compared the cLMA to the Ambu Aura Once, Portex Soft Seal, Boss Systems disposable silicone laryngeal mask, and LMA Unique. We recorded the subjective resistance of the bronchoscope manipulation within the SGA by linear analog score and measured the time to insert the bronchoscope from the proximal end of the SGA to the right upper lobe. We also scored the view of the larynx through the bronchoscope and measured SGA cuff pressures.
Results:
Resistance to bronchoscope manipulation during pediatric flexible bronchoscopy was higher using polyvinyl chloride (PVC) disposable SGAs (Ambu, Unique, and Portex) than the silicone re-usable cLMA (P < 0.0001). The Unique and Ambu laryngeal masks were clinically inferior to the cLMA at all levels of the airway (P < 0.0001). The Portex Soft Seal was not different above the larynx but was significantly statistically inferior at (P < 0.04) and below the larynx (P < 0.006) and inferior overall (P < 0.007). Boss Systems single-use laryngeal mask was as effective as the cLMA.
Conclusion:
In this trial, PVC single-use laryngeal masks were inferior to the silicone cLMA and Boss Systems laryngeal masks for flexible bronchoscopy in children.
Related Concept Videos
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Endoscopic Studies I: Bronchoscopy and Thoracoscopy
Bronchoscopy
Description
Bronchoscopy is a procedure that involves direct visualization of the larynx, trachea, and bronchi for diagnostic and therapeutic purposes. A flexible fiber optic or rigid bronchoscope is used to carry out the procedure. The fiber-optic bronchoscope is more frequently used due to...
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...
Suctioning the Nasopharyngeal Airway
Equipment Required