Corrosion of modular hip prostheses.

E B Mathiesen1, J U Lindgren, G G Blomgren

  • 1Karolinska Institute, Huddinge, Sweden.

The Journal of Bone and Joint Surgery. British Volume
|July 1, 1991
PubMed
Summary

Modular hip prostheses revisions revealed corrosion and metal toxicity. Structural flaws in prosthetic heads, particularly at the head-neck junction, may cause these issues, leading to adverse patient outcomes.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Short-term perioperative parecoxib is not detrimental to shaft fracture healing in a rat model.

Bone & joint research·2019
Same author

Repeated measurements of carotid atherosclerosis and future risk of venous thromboembolism: the Tromsø Study.

Journal of thrombosis and haemostasis : JTH·2017
Same author

Atrial fibrillation is associated with cognitive decline in stroke-free subjects: the Tromsø Study.

European journal of neurology·2017
Same author

2016 Comprehensive Update of the Banff Working Group on Liver Allograft Pathology: Introduction of Antibody-Mediated Rejection.

American journal of transplantation : official journal of the American Society of Transplantation and the American Society of Transplant Surgeons·2016
Same author

C-reactive protein, obesity, and the risk of arterial and venous thrombosis.

Journal of thrombosis and haemostasis : JTH·2016
Same author

Impact of incident myocardial infarction on the risk of venous thromboembolism: the Tromsø Study.

Journal of thrombosis and haemostasis : JTH·2016

Area of Science:

  • Orthopedic Surgery
  • Biomaterials Science
  • Failure Analysis

Background:

  • Modular hip prostheses are widely used in orthopedic surgery.
  • Concerns exist regarding the long-term performance and potential complications of modular implant designs.

Purpose of the Study:

  • To investigate the causes of failure in revised uncemented modular hip prostheses.
  • To identify potential sites of corrosion and material degradation in these implants.

Main Methods:

  • Analysis of nine revised uncemented modular hip prostheses.
  • Surgical findings including tissue examination.
  • Histological and metallographic examination of prosthetic components.

Main Results:

  • Four out of nine prostheses showed macroscopic corrosion at the head-neck junction.
  • Histology revealed extensive necrosis and metal particles, suggesting metal toxicity.
  • Metallography indicated structural imperfections and inhomogeneity in prosthetic heads.

Conclusions:

  • The head-neck junction in modular hip prostheses is a potential site for corrosion.
  • Material defects and corrosion may contribute to adverse tissue reactions and implant failure.
  • Further research into modular interface design and material integrity is warranted.

Related Experiment Videos