Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
Quantification and reduction of peripheral dose from leakage radiation on Siemens Primus accelerators in electron
Collins Yeboah1, Alex Karotki, Dylan Hunt
1Department of Medical Physics, Odette Cancer Centre, Toronto, Ontario, Canada. collins.yeboah@sunnybrook.ca
Abstract:
In this work, leakage radiation from EA200 series electron applicators on Siemens Primus accelerators is quantified, and its penetration ability in water and/or the shielding material Xenolite-NL established. Initially, measurement of leakage from 10 x 10 - 25 x 25 cm2 applicators was performed as a function of height along applicator and of lateral distance from applicator body. Relative to central-axis ionization maximum in solid water, the maximum leakage in air observed with a cylindrical ion chamber with 1 cm solid water buildup cap at a lateral distance of 2 cm from the front and right sidewalls of applicators were 17% and 14%, respectively; these maxima were recorded for 18 MeV electron beams and applicator sizes of >or=20 x 20 cm2. In the patient plane, the applicator leakage gave rise to a broad peripheral dose off-axis distance peak that shifted closer to the field edge as the electron energy increases. The maximum peripheral dose from normally incident primary electron beams at a depth of 1 cm in a water phantom was observed to be equal to 5% of the central-axis dose maximum and as high as 9% for obliquely incident beams with angles of obliquity