Caution on the use of liquid nitrogen traps in stable hydrogen isotope-ratio mass spectrometry
1U.S. Geological Survey, 431 National Center, Reston, Virginia 20192, USA. tbcoplen@usgs.gov
Abstract:
An anomalous stable hydrogen isotopic fractionation of 4 ‰ in gaseous hydrogen has been correlated with the process of adding liquid nitrogen (LN(2)) to top off the dewar of a stainless-steel water trap on a gaseous hydrogen-water platinum equilibration system. Although the cause of this isotopic fractionation is unknown, its effect can be mitigated by (1) increasing the capacity of any dewars so that they do not need to be filled during a daily analytic run, (2) interspersing isotopic reference waters among unknowns, and (3) applying a linear drift correction and linear normalization to isotopic results with a program such as Laboratory Information Management System (LIMS) for Light Stable Isotopes. With adoption of the above guidelines, measurement uncertainty can be substantially improved. For example, the long-term (months to years) δ(2)H reproducibility (1σ standard deviation) of nine local isotopic reference waters analyzed daily improved substantially from about 1 ‰ to 0.58 ‰. This isotopically fractionating mechanism might affect other isotope-ratio mass spectrometers in which LN(2) is used as a moisture trap for gaseous hydrogen.
More Related Videos
Related Concept Videos
¹H NMR of Labile Protons: Deuterium (²H) Substitution
Chemical Shift: Internal References and Solvent Effects
The internal reference compound generally used in NMR spectroscopy is tetramethylsilane (TMS). TMS is preferred because it is chemically inert, soluble in NMR solvents, and easily removable. Also, the highly shielded methyl protons in TMS yield an intense...
Mass Spectrometry: Isotope Effect
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
High-Resolution Mass Spectrometry (HRMS)
Gas Chromatography: Types of Detectors-II