The use of continuous quality-improvement (CQI) processes in the implementation of intelligent infusion pumps in a community teaching hospital is described.
Summary:
After the decision was made to implement intelligent i.v. infusion pumps in a 413-bed, community teaching hospital, drug libraries for use in the safety software had to be created. Before drug libraries could be created, it was necessary to determine the epidemiology of medication use in various clinical care areas. Standardization of medication administration was performed through the CQI process, using practical knowledge of clinicians at the bedside and evidence-based drug safety parameters in the scientific literature. Post-implementation, CQI allowed refinement of clinically important safety limits while minimizing inappropriate, meaningless soft limit alerts on a few select agents. Assigning individual clinical care areas (CCAs) to individual patient care units facilitated customization of drug libraries and identification of specific CCA compliance concerns. Between June 2007 and June 2008, there were seven library updates. These involved drug additions and deletions, customization of individual CCAs, and alterations of limits. Overall compliance with safety software use rose over time, from 33% in November 2006 to over 98% in December 2009. Many potentially clinically significant dosing errors were intercepted by the safety software, prompting edits by end users. Only 4-6% of soft limit alerts resulted in edits.
Conclusion:
Compliance rates for use of infusion pump safety software varied among CCAs over time. Education, auditing, and refinement of drug libraries led to improved compliance in most CCAs.
Related Concept Videos
Drug Accumulation During Multiple Dosing: Intermittent IV Infusions
One-Compartment Model: IV Infusion
The one-compartment model for IV infusion uses mathematical equations to describe the rate of change in drug quantity in the body. At steady-state or infusion equilibrium, the drug input...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Two-Compartment Open Model: IV Infusion
The model illustrates the decrease in plasma drug concentration from the central compartment with a specific equation. It shows that under steady-state conditions, the drug's input rate...
Determination of Multiple Dosing Parameters: Steady-State, Minimum and Maximum Concentrations
Health Information Technology and Healthcare Information System
Health Information Technology, commonly called HIT, integrates advanced information systems and technology in healthcare settings. Its primary functions include: