Design and testing of a new high-accuracy ultraviolet-visible-near-infrared spectrophotometer
Abstract:
A new high-accuracy spectrophotometer has been developed at the National Research Council of Canada to measure regular transmittance factors over the spectral range from 200 to 2500 nm. The most significant feature of this automated single-beam instrument is a highly collimated normal-incidence beam geometry, which eliminates the need for polarization corrections or for an averaging sphere for the calibration of regular-transmittance reference materials. The instrument also possesses a large uniformmeasurement beam that minimizes errors caused by sample nonuniformity. We describe the instrument's design and the testing, optimization, and verification procedures that have been carried out for measurements in the visible and near-infrared regions. Systematic errors that have been determined and corrected for include wavelength shifts, stray light, system drift, and nonlinearity. In the visible and near-infrared regions, the overall photometric accuracy is estimated to be 2.5 and 4.0 parts in 10(4), respectively. The wavelength scale is accurate to within +/-0.1 nm with a reproducibility of +/-0.03 nm over its entire design range from 200 to 2500 nm.
Related Concept Videos
UV–Vis Spectrometers
Ultraviolet and Visible (UV–Vis) Spectroscopy: Overview
Spectrophotometry: Introduction
The essential components of a spectrophotometer include a source of electromagnetic radiation, a slot for placing a material to be analyzed, and a...
UV–Vis Spectroscopy: Beer–Lambert Law
IR Spectrometers
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...