Individualized Stem-positioning in Calcar-guided Short-stem Total Hip Arthroplasty
Published on: February 27, 2018
Biomechanical evaluation of acetabular cup implantation in cementless total hip arthroplasty
Kate A Margalit1, Kei Hayashi, Josh Jackson
1Veterinary Specialty Hospital, San Diego, CA, USA.
Objective:
To report biomechanical properties of the Biologic Fixation System (BFX) acetabular cup impacted into a normal canine pelvis and to compare the effect of implant positioned to and beyond the medial acetabular wall.
Study Design:
In vitro cadaveric study.
Animals:
Hemipelves of mature, large-breed dogs (n=6).
Methods:
For each dog, 1 hemipelvis was reamed to the depth of the acetabular wall (group A) and 1 was reamed an additional 6 mm after penetration of the medial cortex of the acetabulum (group B). The hemipelves were implanted with acetabular cups and loaded in compression through a matching femoral prosthetic component until failure. Specimen stiffness, and failure displacement, load, and energy were determined from load and displacement data and results between groups compared with a paired t-test.
Results:
Mean failure load was greater in group A (3812 ± 391 N) than group B (2924 ± 316 N; P<.014). No other differences (P>.05) were observed between groups. Bone fracture (n=5) and cup displacement (1) occurred in group A whereas in group B there were 3 fractures and 3 cup displacements.
Conclusions:
Although medial placement of the BFX cup affected compressive failure loads, failure loads for both groups exceeded normal physiologic loads.
Clinical Relevance:
Medial positioning of the acetabular cup does not appear to compromise acetabular implant-pelvic stability under normal physiologic loads. Because arthroplasty candidates often have abnormal acetabular architecture, mechanical properties of the cup placed in acetabula without a dorsal rim should be investigated.