Repeatability of measuring corneal subbasal nerve fiber length in individuals with type 2 diabetes
Nathan Efron1, Katie Edwards, Nicola Roper
1Institute of Health and Biomedical Innovation and School of Optometry, Queensland University of Technology, Queensland, Australia. n.efron@qut.edu.au
Purpose:
To analyze the repeatability of measuring nerve fiber length (NFL) from images of the human corneal subbasal nerve plexus using semiautomated software.
Methods:
Images were captured from the corneas of 50 subjects with type 2 diabetes mellitus who showed varying severity of neuropathy, using the Heidelberg Retina Tomograph 3 with Rostock Corneal Module. Semiautomated nerve analysis software was independently used by two observers to determine NFL from images of the subbasal nerve plexus. This procedure was undertaken on two occasions, 3 days apart.
Results:
The intraclass correlation coefficient values were 0.95 (95% confidence intervals: 0.92-0.97) for individual subjects and 0.95 (95% confidence intervals: 0.74-1.00) for observer. Bland-Altman plots of the NFL values indicated a reduced spread of data with lower NFL values. The overall spread of data was less for (a) the observer who was more experienced at analyzing nerve fiber images and (b) the second measurement occasion.
Conclusions:
Semiautomated measurement of NFL in the subbasal nerve fiber layer is highly repeatable. Repeatability can be enhanced by using more experienced observers. It may be possible to markedly improve repeatability when measuring this anatomic structure using fully automated image analysis software.