Related Experiment Video
Updated: Jun 10, 2026

Minimally Invasive Murine Laryngoscopy for Close-Up Imaging of Laryngeal Motion During Breathing and Swallowing
Published on: December 1, 2023
Optimal degree of mouth opening for laryngeal mask airway function during oral surgery
Takuro Sanuki1, Shingo Sugioka, Motoko Hirokane
1Department of Anesthesiology, Osaka Dental University, Chuo-ku, Osaka, Japan. odu9847@yahoo.co.jp
Purpose:
This study was performed to determine the optimal degree of mouth opening in anesthetized patients requiring laryngeal mask airway (LMA) during oral surgery.
Patients And Methods:
A single, experienced LMA user inserted the LMA in 15 patients who were scheduled for elective oral surgery. Oropharyngeal leak pressure, intracuff pressure, and fiberoptic assessment of the LMA position were sequentially documented in 5 mouth conditions-opening of 1.4 (neutral position), 2, 3, 4, and 5 cm-and any resulting ventilatory difficulties were recorded.
Results:
Oropharyngeal leak pressure with the mouth open 4 cm (21.8 ± 3.2 cm H(2)O, P = .025) and 5 cm (27.3 ± 7.2 cm H(2)O, P < .001) was significantly higher than in the neutral position (18.1 ± 1.5 cm H(2)O), as was intracuff pressure (neutral position, 60.0 ± 0 cm H(2)O; 4 cm, 72.6 ± 5.1 cm H(2)O [P < .001]; and 5 cm, 86.9 ± 14.4 cm H(2)O [P < .001]). LMA position, observed by fiberoptic bronchoscopy, was unchanged by mouth opening, being similar in the 5 mouth conditions (P = .999). In addition, ventilation difficulties (abnormal capnograph curves or inadequate tidal volume) occurred in 2 of 15 patients (13%) and 7 of 15 patients (53%) (P < .001) with the mouth opening of 4 and 5 cm, respectively.
Conclusions:
This study showed that a mouth opening over 4 cm led to substantial increases in oropharyngeal leak pressure and intracuff pressure of the LMA, warranting caution, because gastric insufflation, sore throat, and ventilation difficulties may occur. A mouth opening of 3 cm achieves acceptable airway conditions for anesthetized patients requiring LMA.
Related Concept Videos
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...
Cardiopulmonary Resuscitation II: ACLS Airway Management
Suctioning the Nasopharyngeal Airway
Equipment Required
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Tracheostomy Decannulation
Description of the Procedure
Decannulation refers to the permanent removal of the tracheostomy tube, signaling the resolution of the condition that initially necessitated the tracheostomy. The process requires a well-coordinated interplay between...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...