Diffusion tensor MRI of the kidney at 3.0 and 1.5 Tesla
Aki Kido1, Masako Kataoka, Akira Yamamoto
1Department of Diagnostic Imaging and Nuclear Medicine, Kyoto University, Shogoin, Sakyo-ku, Kyoto, Japan. akikido@kuhp.kyoto-u.ac.jp
Background:
Diffusion tensor imaging (DTI) at 3 T provides information on the microstructure and pathophysiology of tissues that is not available from conventional imaging with an advantage of high signal to noise ratio (SNR).
Purpose:
To evaluate the feasibility of DTI of the normal kidney at 3.0 T compared to results obtained at 1.5 T.
Material And Methods:
DTI of the normal kidney of 15 healthy volunteers obtained with 3.0 and 1.5 T scanners using respiration-triggered acquisition was examined. Fractional anisotropy (FA) and apparent diffusion coefficient (ADC) values of both the renal cortex and the medulla and SNRs were measured (b-values 0 and 400 s/mm², diffusion direction of 6). The image quality of FA and ADC maps was also compared subjectively.
Results:
The FA values of the renal cortex were 0.15 ± 0.03 at 3.0 T and 0.14± 0.03 at 1.5 T on average. This difference was not significant. The FA values of the renal medulla were 0.49 ±0.04 at 3.0 T and 0.42 ± 0.05 at 1.5 T. ADC values of the renal cortex were 2.46 x 10⁻³± 0.09 mm<²/s at 3.0 T and 2.20 x 10⁻³±0.11 mm²/s at 1.5 T. The ADC values of the renal medulla were 2.08 x 10⁻³ ± 0.08 mm²/s at 3.0 T and 1.90 x 10⁻³± 0.11 mm²/s at 1.5 T. These FA and ADC values were consistent with previous publications. The difference was significant for the FA value of the medulla (P< 0.01) and ADC values in both cortex and medulla (P < 0.01). The subjective image quality of the FA map with the 3.0 T scanner was significantly superior to that with the 1.5 T scanner (P< 0.01), but not significant for the ADC map (P = 0.18). There was a significant difference in SNR between 3.0 T (48.8 ± 6.6) and 1.5 T images (32.8 ± 5.0).
Conclusion:
The feasibility of renal DTI with a 3.0 T magnet resulting in improved SNR was demonstrated.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies I: Kidney, Ureter, and Bladder Studies
Imaging Studies IV: Magnetic Resonance Imaging
Imaging Studies III: Computed Tomography
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...