In Vitro Evaluation of The Effects Of Er,Cr:YSGG and Diode Lasers Used on Titanium Cylinder
Published on: June 6, 2025
Temperature rise during Er:YAG cavity preparation of primary enamel
Marta Maria Martins Giamatei Contente1, Fabrício Augusto de Lima, Rodrigo Galo
1Department of Pediatric Clinic, Preventive and Social Dentistry, School of Dentistry of Ribeirão Preto, University of São Paulo, Ribeirão Preto, SP, Brazil.
Abstract:
This study aimed to assess in vitro thermal alterations taking place during the Er:YAG laser cavity preparation of primary tooth enamel at different energies and pulse repetition rates. Forty healthy human primary molars were bisected in a mesio-distal direction, thus providing 80 fragments. Two small orifices were made on the dentin surface to which type K thermocouples were attached. The fragments were individually fixed with wax in a cylindrical Plexiglass® abutment and randomly assigned to eight groups, according to the laser parameters (n = 10): G1 - 250 mJ/ 3 Hz, G2 - 250 mJ/ 4 Hz, G3 - 250 mJ/ 6 Hz, G4 - 250 mJ/10 Hz, G5 - 250 mJ/ 15 Hz, G6 - 300 mJ/ 3 Hz, G7 - 300 mJ/ 4 Hz and G8 - 300 mJ/ 6 Hz. An area of 4 mm(2) was delimited. Cavities were done (2 mm long × 2 mm wide × 1 mm thick) using non-contact (12 mm) and focused mode. Temperature values were registered from the start of laser irradiation until the end of cavity preparation. Data were analyzed by one-way ANOVA and Tukey test (p ≤ 0.05). Groups G1, G2, G6, and G7 were statistically similar and furnished the lowest mean values of temperature rise. The set 250 mJ/10 and 15 Hz yielded the highest temperature values. The sets 250 and 300 mJ and 6 Hz provided temperatures with mean values below the acceptable critical value, suggesting that these parameters ablate the primary tooth enamel. Moreover, the temperature elevation was directly related to the increase in the employed pulse repetition rates. In addition, there was no direct correlation between temperature rise and energy density. Therefore, it is important to use a lower pulse frequency, such as 300 mJ and 6 Hz, during cavity preparation in pediatric patients.
Related Concept Videos
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Preparation of Samples for Electron Microscopy