A novel external counterpulsation system for coronary artery disease and heart failure: pilot studies and initial
Tsuyoshi Shimizu1, Shunei Kyo, Kazuhito Imanaka
1Department of Therapeutic Strategy for Heart Failure, Graduate School of Medicine, The University of Tokyo, 7-3-1 Hongo, Bunkyo-ku, Tokyo, 113-8655, Japan. tshimizu-cvs@umin.ac.jp
Insights
A novel external counterpulsation (ECP) system, Compact CP, demonstrated safety and effectiveness in pilot studies for coronary artery disease and heart failure. This innovative device shows promise for improving patient outcomes with no adverse effects observed.
Area of Science:
- Cardiovascular Medicine
- Biomedical Engineering
Background:
- External counterpulsation (ECP) is a noninvasive treatment for cardiovascular conditions but has limitations.
- Existing ECP systems can be bulky and impactful due to cuff inflation and compressor size.
Purpose of the Study:
- To evaluate a novel ECP system, Compact CP, featuring a double-lumen cuff.
- To assess the safety, efficacy, and comfort of the Compact CP system in healthy volunteers and patients with cardiovascular diseases.
Main Methods:
- The Compact CP system utilizes a double-lumen cuff: a contact cuff with constant pressure and a main cuff synchronized to the cardiac cycle.
- Four pilot studies were conducted on 39 healthy volunteers, alongside initial clinical application in three patients.
Main Results:
- The system achieved significant diastolic augmentation and systolic unloading in pilot studies.
- A satisfactory diastolic/systolic pressure ratio (1.00 ± 0.06) and high comfort were noted at 40 kPa driving pressure.
- Clinical improvements were observed in patients with chronic refractory angina and heart failure, with no adverse effects.
Conclusions:
- The novel Compact CP ECP system is safe and effective for treating coronary artery disease and heart failure.
- The system demonstrates promising results, warranting further clinical investigation.
Abstract:
External counterpulsation (ECP) is a beneficial and noninvasive treatment for coronary artery disease or heart failure; however, it still has a lot of limitations. We used a novel ECP system, Compact CP, the main feature of which is the double-lumen cuff that reduces the impact of cuff inflation and the size of the air compressor. The first lumen was a contact cuff that was attached to the legs with a constant pressure (8 kPa). The second lumen was a main cuff that was inflated and deflated with a driving pressure and synchronized to the cardiac cycle. In this report, we describe the results of four pilot studies in a total number of 39 healthy volunteers and initial clinical experiences of this system in three patients. The pilot studies demonstrated that the ECP system provided significant diastolic augmentation and systolic unloading. It also achieved a satisfactory diastolic/systolic pressure ratio (1.00 ± 0.06) with a high comfort level at a driving pressure of 40 kPa. Higher pressure (50-70 kPa) increased the assist performance but decreased the comfort level. ECP was also applied with a patient with chronic refractory angina and two patients with postoperative heart failure following cardiac surgery. The clinical conditions improved. No adverse effect was observed. Our novel ECP system is safe, effective, and promising in the treatment of coronary artery disease or heart failure. Further clinical investigations are needed to support the significance of this system.