Characteristics of patients and implantable defibrillators associated with failure to sense device alert systems
Matthew T Bennett1, Charles R Kerr, Edna Hahn
1Division of Cardiology, St Paul's Hospital, University of British Columbia, ICD Clinic, 1081 Burrard St., Vancouver, BC, Canada.
Aims:
In the era of increasing implantable cardioverter defibrillator (ICD) complexity, the ICD patient alert is deemed to be an important feature in the early detection of ICD system malfunction and is either an audible or a vibratory alert. We sought to evaluate the patient's ability to detect these ICD alerts in the device clinic setting as a surrogate endpoint of clinical utility.
Methods And Results:
From 1 November 2006 to 31 March 2008, 563 patients with an ICD equipped with either an audible patient alert (APA, Medtronic and Guidant; n = 485) or a vibratory monitoring alert ([VMA, St Jude Medical; n = 78) had their alarm demonstrated in the quiet clinic setting. The ability to recognize the alert was analysed and then stratified by gender, age, manufacturer, type of alert, and pocket location. The average patient age was 63.3 (± 13.6) years and 82.8% of patients were male. Implantable cardioverter defibrillator manufacturers were Medtronic (n = 464), Boston Scientific (n = 21), and SJM (n = 78). The APA was heard in 86.0% of patients. This was less likely in patients who were older, male, and where the device was placed in the submuscular position. Every patient with a VMA sensed their alert.
Conclusion:
In the current ICD alert technology, the ability to sense the ICD alert in the device clinic appears to be higher for the VMA than for the APA. In particular, older patients and male patients are less likely to sense the APA.
Related Concept Videos
Cardiopulmonary Resuscitation III: AED Use
Cardiomyopathy II: Dilated Cardiomyopathy
Cardiomyopathy V: Interprofessional Care
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Dysrhythmias IV: Characteristics of Bradyarrhythmias
Mechanism of Cardiac Arrhythmias